Abstract #10270 Published in IGR 6-1

Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness

Miglior S; Albe E; Guareschi M; Mandelli G; Gomarasca S; Orzalesi N
British Journal of Ophthalmology 2004; 88: 174-177


PURPOSE: To assess reproducibility of central corneal thickness (CCT) measurement by means of ultrasonic pachymetry. METHODS: Fifty-one volunteers underwent three sessions of CCT measurements, each consisting of three CCT measurements, performed by each of three different observers. Intra- and interobserver reproducibility was calculated by means of intraclass correlation coefficient (ICC). The expected range of variability between two independent evaluations was calculated using scatter plots of each test-retest difference against their mean. The standard deviation of the mean differences in the test-retest scores was used to describe the differences in the score spread. RESULTS: The ICC ranges of the intra- and interobserver evaluations were 0.95-0.97 and 0.89-0.95, respectively; the expected variability was ≤ ± 1% and 2%, respectively (95% confidence interval). CONCLUSION: Measurement of CCT by means of ultrasonic pachymetry is highly reproducible.