advertisement

Topcon

Abstract #108554 Published in IGR 23-4

Prediction model of contact forces and IOP during digital palpation of porcine eyes

Zhang Q; Enikov ET
Journal of the mechanical behavior of biomedical materials 2023; 142: 105864


Frequent intraocular pressure (IOP) measurements are desirable in the diagnosis and management of glaucoma. Most current tonometers utilize some form of corneal deformation to estimate the IOP, since trans-scleral tonometry suffers from loss of sensitivity. Tran-scleral and trans-palpebral tonometry, however, offer a pathway towards a non-invasive home tonometry. This article presents a mathematical model capturing the relationship between the IOP and the displacements imposed onto the sclera by externally applied forces. Similar to manual digital palpation tonometry, trans-scleral mechanical palpation makes use of two force probes that are advanced in a specific order and distance. Data from the applied forces and displacements, along with concurrent measurements of IOP is used to produce a phenomenological mathematical model. The experiments were carried out on enucleated porcine eyes. Two models are presented. Model 1 predicts IOP vs forces and displacements, while Model 2 predicts the baseline IOP (prior to applying the forces) as a function of the measured forces and displacements. The proposed models result in IOP errors of 1.65 mmHG and 0.82 mmHg, respectively. Model parameters were extracted using least-squares-based system identification methods. The results show that the proposed models can be used to estimate the baseline IOP with accuracy of ±1 mmHg over a pressure range of 10-35 mmHg, solely from measurement of tactile forces and displacements.

Advanced Micro- and Nanosystems Laboratory, Aerospace and Mechanical Engineering Department, University of Arizona, 1130 N Mountain Ave., Tucson, AZ 85721-0119, USA.

Full article

Classification:

15 Miscellaneous



Issue 23-4

Change Issue


advertisement

Oculus