PURPOSE: To report post-operative course and complications in patients who underwent Ex-Press glaucoma shunt placement. DESIGN: Small case series. METHODS: Four patients were referred for management following Ex-Press shunt implantation. Patients underwent complete ophthalmologic evaluation and were appropriately managed. RESULTS: Of the four patients, two had inadequate control of intraocular pressure related to bleb failure caused by subconjunctival scar tissue formation. One patient experienced suprachoroidal hemorrhage 5 days after shunt placement followed a year later with conjunctival erosion and shunt rim exposure. Endophthalmitis from conjunctival erosion over the Ex-Press shunt rim was observed in one patient. CONCLUSION: Patients with subconjunctival implantation of the Ex-Press shunt should be monitored closely for possible conjunctival erosion that can lead to endophthalmitis and failure from fibrosis.
Dr. R.M. Stewart, Department of Ophthalmology, Tulane School of Medicine, Tulane University, New Orleans, LA, USA
12.8.2 With tube implant or other drainage devices (Part of: 12 Surgical treatment > 12.8 Filtering surgery)