Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness

Iliev ME; Goldblum D; Katsoulis K; Amstutz C; Frueh B
British Journal of Ophthalmology 2006; 90: 833-835


BACKGROUND/AIMS: Rebound tonometry (RT) is performed without anaesthesia with a hand held device. The primary aim was to compare RT with Goldmann applanation tonometry (GAT) and to correlate with central corneal thickness (CCT). The secondary aim was to prove tolerability and practicability of RT under 'study conditions' and 'routine practice conditions'. METHODS: In group 1 (52 eyes/28 patients), all measurements were taken by the same physician, in the same room and order: non-contact optical pachymetry, RT, slit lamp inspection, GAT. Patients were questioned about discomfort or pain. In group 2 (49 eyes/27 patients), tonometry was performed by three other physicians during routine examinations. RESULTS: RT was well tolerated and safe. Intraocular pressure (IOP) ranged between 6 mmHg and 48 mmHg. No different trends were found between the groups. RT tended to give slightly higher readings: n = 101, mean difference 1.0 (SD 2.17) mmHg; 84.1% of RT readings within plus or minus 3 mmHg of GAT; 95% confidence interval in the Bland-Altman analysis -3.2 mmHg to +5.2 mmHg. Both RT and GAT showed a weak positive correlation with CCT (r2 0.028 and 0.025, respectively). CONCLUSIONS: RT can be considered a reliable alternative for clinical screening and in cases where positioning of the head at the slit lamp is impossible or topical preparations are to be avoided.

Dr. M.E. Iliev, Department of Opthalmology, University of Bern, Inselspital, 3010 Bern, Switzerland. milko.iliev@insel.ch


Classification:

6.1 Intraocular pressure measurement; factors affecting IOP (Part of: 6 Clinical examination methods)



Issue 8-3

Change Issue


advertisement

Oculus