Abstract #14821 Published in IGR 8-4

Comparative measurements of central corneal thickness with two ultrasound pachymeters

Schmidt E; Schwanitz K; Bohm AG; Pillunat LE
Ophthalmologe 2006; 103: 688-692


BACKGROUND: The aim of this study was to compare two different ultrasound pachymeters based on measurements of central corneal thickness (CCT). METHODS: CCT of 1,070 eyes of 535 patients (glaucoma and glaucoma suspects) was determined using the ultrasound pachymeter IOPac (Heidelberg Engineering, Germany) and SP-3000 (Tomey Corporation, Japan). With the IOPac 8 measurements are averaged by the instrument and with the SP-3000 200 measurements, respectively. For statistical purposes intraclass correlation coefficients (ICC) and Bland-Altman analysis were calculated. RESULTS: The mean CCT was 561.3 ± 35.6 μm (IOPac) and 561.8±35.6 μm (SP-3000). The mean difference between both pachymeters was 0 ± 8 μm. ICC was 0.9746 (95% confidence interval: 0.9714-0.9774). CONCLUSIONS: The use of the two instruments tested has no influence on the determination of CCT. The measured CCT of glaucoma patients or suspects were slightly above the CCT of normals of other studies. LA: German

Dr. E. Schmidt, Augenklinik, Universitat Dresden, Fetscherstrasse 74, 01307 Dresden, Germany