Comparison of the Icare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists

Abraham LM; Epasinghe NC; Selva D; Casson R
Eye 2008; 22: 503-506


PURPOSE: To assess the agreement between Icare® rebound tonometer and Goldmann applanation tonometer in the hands of experienced and inexperienced tonometrists. PATIENTS AND METHODS: Two tonometrists, experienced with both Goldmann applanation tonometry (GAT) and ICare® Tonometry (ICT) measured intraocular pressure (IOP), in a masked fashion, in 100 patients. In another series of 58 patients, ICT was performed by an inexperienced tonometrist and GAT by an experienced tonometrist. RESULTS: In approximately 80% of patients, the difference in IOP between GAT and ICT was ≤ 2 mmHg in group 1 and ≤ 3 mmHg in group 2. The 95% limits of agreement were -4.0-4.4 mmHg in group 1 and -6.0-5.0 mmHg in group 2. CONCLUSION: ICT compares reasonably with GAT, in both experienced and inexperienced hands. Its ease of use, portability, and sterility make it an attractive tonometer. Its degree of accuracy in inexperienced hands would make it a useful instrument for health care workers with limited ophthalmic experience.

Dr. L.M. Abraham, Department of Ophthalmology and Visual Sciences, South Australian Institute of Ophthalmology, Royal Adelaide Hospital, Adelaide, South Australia, Australia


Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)



Issue 10-2