advertisement

Topcon

Abstract #21593 Published in IGR 10-3

Tono-Pen XL tonometry during application of a suction ring in rabbits

Charisis SK; Ginis HS; Kounis GA; Tsilimbaris MK
BMC Ophthalmology 2008; 8: 14


BACKGROUND: The purpose of this study is to evaluate the use of Tono-Pen XL in measuring IOP during the application of a suction ring in rabbit eyes with manometrically controlled IOP. METHODS: Tono-Pen XL was calibrated against direct manometry in 10 rabbit eyes. A suction ring was then applied in 4 rabbit eyes and the IOP was determined manometrically during suction ring application at 350 mmHg vacuum pressure. Finally, in 6 catheterized rabbit eyes the IOP was measured with Tono-Pen XL during suction ring application at suction vacuum from 350 to 650 mmHg, while keeping actual IOP stable at 30 mmHg and 60 mmHg. RESULTS: Linear regression analysis revealed that the Tono-pen XL was reliable for IOPs between 10 and 70 mmHg (R2 = 0.9855). Direct manometry during suction ring application showed no statistically significant variation of Tono-Pen XL readings when the incanulation manometry intraocular pressure changed from 30 mmHg to 60 mmHg and no statistically significant correlation between suction vacuum and IOP measurements. CONCLUSION: Tono-Pen XL measurements are unreliable during the application of a suction ring on living rabbit eyes even when the actual IOP is forced to be within the validated range of Tono-Pen XL measurements. This inaccuracy is probably related to altered corneal and scleral geometry and stress.

Dr. S.K. Charisis, University Hospital of Heraklion, Eye Clinic, Heraklion, Crete, Greece. scharisis@gmail.com


Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)
5.3 Other (Part of: 5 Experimental glaucoma; animal models)



Issue 10-3

Change Issue


advertisement

Oculus