advertisement

Topcon

Abstract #23535 Published in IGR 11-2

Effect of image quality on tissue thickness measurements obtained with spectral domain-optical coherence tomography

Balasubramanian M; Bowd C; Vizzeri G; Weinreb RN; Zangwill LM
Optics express 2009; 17: 4019-4036


The purpose of this study was to investigate the effect of image quality on retinal nerve fiber layer (RNFL) and retinal thickness measurements obtained using three commercially available spectral domain-optical coherence tomographers (SD-OCT). Subjectively determined good, medium and poor quality images were obtained from four healthy and one glaucoma suspect eyes. RNFL and retinal thickness measurements were compared as a function of image quality. Results indicate that when image quality is within the range specified as acceptable by SD-OCT manufacturers, RNFL and retinal thickness measurements are comparable.

M. Balasubramanian. Hamilton Glaucoma Center, Department of Ophthalmology, University of California, San Diego, La Jolla CA. 92093-0946 , USA.


Classification:

6.9.2.2 Posterior (Part of: 6 Clinical examination methods > 6.9 Computerized image analysis > 6.9.2 Optical coherence tomography)
2.13 Retina and retinal nerve fibre layer (Part of: 2 Anatomical structures in glaucoma)



Issue 11-2

Change Issue


advertisement

Oculus