Abstract #47992 Published in IGR 13-4

Aligning scan acquisition circles in optical coherence tomography images of the retinal nerve fibre layer

Zhu H; Crabb DP; Schlottmann PG; Wollstein G; Garway-Heath DF
IEEE Transactions on Medical Imaging 2011; 30: 1228-1238


Optical coherence tomography (OCT) is widely used in the assessment of retinal nerve fibre layer thickness (RNFLT) in glaucoma. Images are typically acquired with a circular scan around the optic nerve head. Accurate registration of OCT scans is essential for measurement reproducibility and longitudinal examination. This study developed and evaluated a special image registration algorithm to align the location of the OCT scan circles to the vessel features in the retina using probabilistic modelling that was optimised by an expectation-maximization algorithm. Evaluation of the method on 18 patients undergoing large number of scans indicated improved data acquisition and better reproducibility of measured RNFLT when scanning circles were closely matched. The proposed method enables clinicians to consider the RNFLT measurement and its scan circle location on the retina in tandem, reducing RNFLT measurement variability and assisting detection of real change of RNFLT in the longitudinal assessment of glaucoma.

H. Zhu. Department of Optometry and Visual Science, City University London, London, UK. Email: haogangzhu@gmail.com


Classification:

6.9.2 Optical coherence tomography (Part of: 6 Clinical examination methods > 6.9 Computerized image analysis)
6.9.2.2 Posterior (Part of: 6 Clinical examination methods > 6.9 Computerized image analysis > 6.9.2 Optical coherence tomography)