Comparison of dynamic contour tonometry, Goldmann applanation Tonometry and tonopen for measuring intraocular pressure in normal tension glaucoma

Kurtz S; Soiberman U; Shemesh G
Harefuah 2013; 152: 643-646, 689


BACKGROUND: Normal tension glaucoma (NTG) is a chronic progressive optic neuropathy, characterized by relatively normal intraocular pressure (IOP) measurements. This study utilized three modalities in the measurement of lOP: dynamic contour tonometry (Pascal-DCT), Goldmann applanation tonometry (GAT) and Tonopen XL. OBJECTIVES: To compare the lOP measurements of the three devices in eyes with NTG. METHODS: This retrospective study of thirty-three NTG patients (66 eyes) who underwent lOP measurements using the three devices was approved by a local ethics committee. The lOP measurements were stratified according to central corneal thickness (CCT) and corneal curvature (CC). RESULTS: The mean DCT IOP was higher (16.22 +/- 3.04 mmHg) than the mean GAT IOP (13.14 +/- 2.83 mmHg, P < 0.0042). The mean Tonopen XL lOP was lower than the mean GAT IOP (12.76 +/- 3.11 mmHg, P < 0.001). GAT and DCT measurements were significantly infLuenced by corneal curvature (P = 0.004 and P < 0.0001, respectiveLy). CONCLUSIONS: lOPs measured by GAT or Tonopen XL were consistently lower than by DCT. CC significantly influenced GAT and DCT results.


Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)
9.2.4 Normal pressure glaucoma (Part of: 9 Clinical forms of glaucomas > 9.2 Primary open angle glaucomas)