Comparison between Goldmann, Icare Pro and Corvis ST tonometry

Bañeros-Rojas P; Martinez De La Casa JM; Arribas-Pardo P; Berrozpe-Villabona C; Toro-Utrera P; García-Feijoó J
Archivos de la Sociedad EspaƱola de Oftalmologia 2014; 89: 260-264


PURPOSE: To compare intraocular pressure (IOP) between the new non-contact tonometer Corvis ST (CST), the Goldmann applanation tonometry (GAT) and Icare Pro rebound tonometer (PRO). METHODS: A total of 178 eyes of 178 healthy subjects were selected for the study. Measurements of IOP were made in a random order with GAT, PRO and CST. Central corneal thickness (CCT) was determined by ultrasound pachymetry. The mean of three valid measurements of each variable was used in the statistical analysis. The relationship between the tonometers was established using Bland-Altman plots. RESULTS: Mean IOP was 15.5 ± 2.8 mmHg for GAT, 15.4 ± 2.8 mmHg for CST, and 14.6 ± 2.3 mmHg for PRO. The mean differences between pairs of tonometers were: GAT-PRO=0.9 ± 1.7 mmHg (P<.001), GAT-CST: 0.1 ± 2.2 (P=.398), and PRO-CST: -0.8 ±-0.7 mmHg, p<0.001. A positive relationship was detected between CCT and the three tonometers: GAT: r=0.325, P<.001; PRO: r = 0.385, P<.001, and CST: r = 0.428; P<.001. CONCLUSIONS: The differences found between PRO and GAT were significantly higher than those found between CST and GAT, which showed non-significant differences. The measurements of the three tonometers were affected by the CCT.

Full article

Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)