Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects

Güler M; Bilak S; Bilgin B; Simşek A; Capkin M; Hakim Reyhan A
Journal of Glaucoma 2015; 24: 613-618


PURPOSE: To compare intraocular pressure (IOP) measurements obtained by Icare PRO rebound tonometer (RT), Tomey FT-1000 noncontact tonometer (NCT), and Goldmann applanation tonometer (GAT) in healthy subjects, and to investigate the influence of central corneal thickness (CCT) on IOP measurements. METHODS: A total of 132 eyes of 66 healthy subjects were included in the study. All IOP measurements were applied in a sitting position and always in the same order (RT, NCT, and GAT). CCT was measured by noncontact anterior segment optical coherence tomography. RESULTS: The mean CCT was 523.5±31.2 μm. The mean IOPs with GAT, NCT, and RT measurements were 14.56±2.77, 13.42±2.99, 14.18±2.55 mm Hg, respectively. A significantly good agreement was found between RT and GAT measurements (P=0.515). The mean difference between RT and GAT measurements was -0.38 mm Hg. An underestimation was found in NCT measurements compared with GAT ones as defined from paired comparisons (P=0.003). The mean difference between NCT and GAT measurements was -1.14 mm Hg. There was no significant difference between NCT and RT readings (P=0.069). The mean difference between NCT and RT measurements was -0.76 mm Hg. There was a significant correlation between GAT-NCT (r=0.740, P<0.001), GAT-RT (r=0.743, P<0.001), NCT-RT (r=0.729, P<0.001) measurement methods. The increment in measured IOP for a 10 μm increase in CCT was 0.16, 0.33, and 0.18 mm Hg for the GAT, NCT, and RT, respectively. CONCLUSIONS: The IOP readings with the new Icare PRO RT showed good correlation with those obtained by GAT. Tomey FT-1000 NCT underestimated the IOP compared with GAT.

*Department of Ophthalmology, School of Medicine, Adiyaman University †Department of Ophthalmology, Adiyaman University Education and Research Hospital, Adiyaman, Turkey.

Full article

Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)



Issue 16-3

Change Issue


advertisement

Oculus