PURPOSE: To report the outcomes of toric intraocular lens implantation in patients with glaucoma and corneal astigmatism. METHODS: One hundred twenty-six eyes of 87 patients with glaucoma and corneal astigmatism that underwent cataract surgery with an AcrySof toric intraocular lens (Alcon Laboratories, Inc., Fort Worth, TX) implant were selected for this single-center, retrospective case series. Corrected distance visual acuity, intraocular pressure, and refractive astigmatism were measured in each eye preoperatively and postoperatively. Uncorrected distance visual acuity and toric alignment were measured postoperatively. RESULTS: The uncorrected distance visual acuity was 0.04 ± 0.08 logMAR (20/22 Snellen) for all eyes. Ninety-eight percent of all eyes achieved an uncorrected distance visual acuity of 20/40 or better, with 76% achieving 20/25 or better and 47% achieving 20/20. The corrected distance visual acuity for all eyes was 0.01 ± 0.03 logMAR (20/20.5 Snellen) postoperatively. The refractive cylinder improved from 1.47 ± 1.10 diopters preoperatively to 0.31 ± 0.37 diopters postoperatively. The residual refractive cylinder was 1.00 diopter or less in 97% of eyes, 0.75 diopters or less in 90% of eyes, and 0.50 diopters or less in 83% of eyes. Mean misalignment was 4.4° ± 5.1°. Intraocular pressure decreased by a mean of 2.3 ± 3.3 mm Hg following the surgery. CONCLUSION: Toric intraocular lenses can reliably reduce astigmatism and improve uncorrected vision in eyes with cataract and glaucoma.
Full article
12.12.3 Phacoemulsification (Part of: 12 Surgical treatment > 12.12 Cataract extraction)