Fluid Dynamics of a Novel Micro-Fistula Implant for the Surgical Treatment of Glaucoma

Sheybani A; Reitsamer H; Ahmed II
Investigative Ophthalmology and Visual Science 2015; 56: 4789-4795


PURPOSE: The purpose of this study was to describe the fluidics of a novel non-valved glaucoma implant designed to prevent hypotony and compare the fluidics of this device with two commonly used non-valved glaucoma devices. METHODS: The XEN 45 micro-fistula implant was designed to limit hypotony by virtue of its length and width according to the Hagen-Poiseuille equation. Flow testing was performed using a syringe pump and pressure transducer at multiple flow rates. The pressure differentials across the XEN implant, the Ex-Press implant, and 10 mm of silicone tubing from a Baerveldt implant at a physiologic flow rate (2.5 μL/min) were extrapolated. RESULTS: The XEN 45 achieved a steady-state pressure calculated at 7.56 mm Hg at 2.5 μL/min. At the same flow rate, the Ex-Press device and Baerveldt tubing reached steady-state pressures of 0.09 and 0.01 mm Hg, respectively. CONCLUSIONS: Under flow testing, the XEN micro-fistula implant was able to maintain backpressure above numerical hypotony levels without the use of complex valve systems. This is due to the XEN implant's design, derived from the principles that dictate Newtonian fluids.

Department of Ophthalmology and Visual Sciences, Washington University School of Medicine, St. Louis, Missouri, United States.

Full article

Classification:

12.8.2 With tube implant or other drainage devices (Part of: 12 Surgical treatment > 12.8 Filtering surgery)