PURPOSE: We investigated correlations between short-term and long-term intraocular pressure (IOP) fluctuations. METHODS: We examined 50 eyes of glaucoma patients who were followed for >2 years. We measured short-term IOP fluctuation using a Triggerfish(®) contact lens sensor (CLS). The short-term IOP fluctuation (mVeq) was defined as the difference between the maximum value and the minimum value measured during the 24-hour course with CLS. The long-term IOP fluctuation was defined by four parameters: 1) the mean IOP (mmHg) determined during follow-up; 2) the IOP difference, which was defined as the difference between the maximum IOP and the minimum IOP; 3) the standard deviation of IOP; and 4) the peak IOP, which was defined as the maximum IOP. Correlations between these parameters and the short-term IOP fluctuation were examined. RESULTS: The mean follow-up period was 5.4 years. The average IOP was 15.0±4.0 mmHg. The range of short-term IOP fluctuation identified with CLS was significantly correlated with all the four long-term IOP fluctuation parameters. CONCLUSION: Short-term IOP fluctuations were found to be associated with long-term IOP fluctuations. Examination of 24-hour IOP fluctuations with the CLS might be useful for predicting the long-term IOP fluctuation.