Visual field examination method using virtual reality glasses compared with the Humphrey perimeter

Tsapakis S; Papaconstantinou D; Diagourtas A; Droutsas K; Andreanos K; Moschos MM; Brouzas D
Clinical Ophthalmology 2017; 11: 1431-1443


PURPOSE: To present a visual field examination method using virtual reality glasses and evaluate the reliability of the method by comparing the results with those of the Humphrey perimeter. MATERIALS AND METHODS: Virtual reality glasses, a smartphone with a 6 inch display, and software that implements a fast-threshold 3 dB step staircase algorithm for the central 24° of visual field (52 points) were used to test 20 eyes of 10 patients, who were tested in a random and consecutive order as they appeared in our glaucoma department. The results were compared with those obtained from the same patients using the Humphrey perimeter. RESULTS: High correlation coefficient (r=0.808, P<0.0001) was found between the virtual reality visual field test and the Humphrey perimeter visual field. CONCLUSION: Visual field examination results using virtual reality glasses have a high correlation with the Humphrey perimeter allowing the method to be suitable for probable clinical use.

1st Department of Ophthalmology, National and Kapodistrian University of Athens, Athens, Greece.

Full article

Classification:

6.6.2 Automated (Part of: 6 Clinical examination methods > 6.6 Visual field examination and other visual function tests)
6.6.3 Special methods (e.g. color, contrast, SWAP etc.) (Part of: 6 Clinical examination methods > 6.6 Visual field examination and other visual function tests)