Abstract #80816 Published in IGR 20-3

The Outcomes of a Comprehensive Program for Maintenance of Goldmann Applanation Tonometer

Choudhari NS; Richhariya A; Wadke V; Wadke V; Deshmukh SP; George R; Senthil S; Chandra Sekhar G
Journal of Glaucoma 2019; 28: 507-511


PRECIS: We checked 190 tonometers every month and repaired faulty ones. Calibration error (CE) frequency reduced from 23% to 0.6% at 1 year. Tonometers needing one or >1 CE repair differed in survival but not in age. PURPOSE: The purpose of this study was to report the outcomes of a comprehensive program to maintain calibration status of the Goldmann applanation tonometer. METHODS: This prospective cohort study was carried out at 2 tertiary eye care referral centers. We included 190 slit-lamp mounted Goldmann applanation tonometers (Model AT 900 C/M). Health care providers (error checking and reporting) and clinical engineers (maintenance) participated. The team carried out CE check once a month, and repair of faulty tonometers, if any, within 24 hours. Failure of tonometer was defined as development of unacceptable CE beyond the third repair. The main outcome measures were the frequency of CE and survival function of the tonometer over 1 year. RESULTS: The median age of the tonometers was 10.7 (range, 0.2 to 25.1) years. The total number of repairs was 86. The proportion (95% confidence interval) of faulty tonometers reduced from 23.1% (17.7, 29.6) in the first month to 0.6% (0.1, 3.3) at 1 year (P<0.01). The median age of the tonometer did not differ between those needing (n=63, 9.4 y) and not needing (n=127, 10.7 y; P=0.24) repair. All tonometers requiring 1 CE repair (n=49, 25.7%) survived until 1 year. The survival of tonometers requiring >1 CE repair (n=14, 7.3%) was 40% at 1 year. CONCLUSIONS: Our in-house program maintained 92.6% tonometers error free. Number of repairs rather than age determined the need for replacement/sending back the tonometer to the manufacturer. Our simple and easy to follow maintenance program has the potential for wide application.

V S T Glaucoma Center, Dr. Kallam Anji Reddy Campus.

Full article

Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)



Issue 20-3

Change Issue


advertisement

Topcon