Clinical validation of , an automated optic nerve head analysis software

Singh D; Gunasekaran S; Hada M; Gogia V
Indian Journal of Ophthalmology 2019; 67: 1089-1094


PURPOSE: To clinically validate a new automated glaucoma diagnosis software RIA-G. METHODS: A double-blinded study was conducted where 229 valid random fundus images were evaluated independently by RIA-G and three expert ophthalmologists. Optic nerve head parameters [vertical and horizontal cup-disc ratio (CDR) and neuroretinal rim (NRR) changes] were quantified. Disc damage likelihood scale (DDLS) staging and presence of glaucoma were noted. The software output was compared with consensus values of ophthalmologists. RESULTS: Mean difference between the vertical CDR output by RIA-G and the ophthalmologists was - 0.004 ± 0.1. Good agreement and strong correlation existed between the two [interclass correlation coefficient (ICC) 0.79; r = 0.77, P < 0.005]. Mean difference for horizontal CDR was - 0.07 ± 0.13 with a moderate to strong agreement and correlation (ICC 0.48; r = 0.61, P < 0.05). Experts and RIA-G found a violation of the inferior-superior NRR in 47 and 54 images, respectively (Cohen's kappa = 0.56 ± 0.07). RIA-G accurately detected DDLS in 66.2% cases, while in 93.8% cases, output was within ± 1 stage (ICC 0.51). Sensitivity and specificity of RIA-G to diagnose glaucomatous neuropathy were 82.3% and 91.8%, respectively. Overall agreement between RIA-G and experts for glaucoma diagnosis was good (Cohen's kappa = 0.62 ± 0.07). Overall accuracy of RIA-G to detect glaucomatous neuropathy was 90.3%. A detection error rate of 5% was noted. CONCLUSION: RIA-G showed good agreement with the experts and proved to be a reliable software for detecting glaucomatous optic neuropathy. The ability to quantify optic nerve head parameters from simple fundus photographs will prove particularly useful in glaucoma screening, where no direct patient-doctor contact is established.

Noble Eye Care; Narayana Superspecialty Hospital, Gurugram, Haryana, India.

Full article

Classification:

6.9.5 Other (Part of: 6 Clinical examination methods > 6.9 Computerized image analysis)
6.8.2 Posterior segment (Part of: 6 Clinical examination methods > 6.8 Photography)



Issue 20-3

Change Issue


advertisement

Oculus