BMC Ophthalmology 2021; 21: 273


BACKGROUND: PreserFlo® MicroShunt (PM) (also known as InnFocus® MicroShunt) is a subconjunctival stent implanted ab externo via a minimally invasive surgical procedure. The current indication is progressive, mild to moderate, open angle glaucoma uncontrolled on topical medications. According to the literature, adverse events are rare, mild and transient. CASE PRESENTATION: Two cases of stand-alone PreserFlo MicroShunt® implantation in patients with uncontrolled open-angle glaucoma are reported. Exposure occurred 7 days and 3 months respectively after implantation. These cases shared common features including preexisting blepharitis and the lack of a Tenon's flap. In both cases, removal of the device was required after several attempts at repair. CONCLUSIONS: PreserFlo MicroShunt® exposure is a potentially vision-threatening complication because of the risk of endophthalmitis. Potential risk factors include the absence of a Tenon's flap and pre-existing ocular surface inflammation. Ocular surface inflammation should be detected and treated prior to PM implantation. If a deficiency in Tenon's capsule is noted intraoperatively, close monitoring should be performed because of the higher risk of PM exposure.

Department of Ophthalmology III, Quinze-Vingts National Ophthalmology Hospital, 28 rue de Charenton, 75012, Paris, France.

Full article