advertisement
The main reason for the failure of glaucoma drainage devices is the chronic inflammatory reaction caused by the poor biocompatibility of biomaterials, which leads to the migration and proliferation of scleral fibroblasts surrounding the devices thus forming the encapsulation that will block the aqueous humor drainage channel. In order to inhibit the reaction of scleral fibroblasts after glaucoma drainage device implantation, this study designed and fabricated a 5-fluorouracil (5-FU) loaded chitosan microtube (CMT) for glaucoma aqueous humor drainage. The CMTs were made by template based adsorption-precipitation method using chitosan with excellent biocompatibility as raw material due to its characteristic of pH dependent solubility. The physical properties of CMTs were investigated. The and biocompatibilities were studied as well. The developed drainage device is expected to possess the dual function of reducing intraocular pressure and inhibiting excessive fibrosis after glaucoma drainage device implantation, thereby providing a novel way for the research of glaucoma drainage devices.
School of Ophthalmology & Optometry, Eye Hospital, Wenzhou Medical University, Wenzhou, China.
Full article12.8.2 With tube implant or other drainage devices (Part of: 12 Surgical treatment > 12.8 Filtering surgery)
12.8.10 Woundhealing antifibrosis (Part of: 12 Surgical treatment > 12.8 Filtering surgery)