PURPOSE: To analyse the morphological evolution of filtering blebs with anterior-segment OCT (AS-OCT) and its correlation with intraocular pressure after ab externo SIBS microshunt implantation with mitomycin C (MMC) during a 3-month follow-up period. METHODS: Twenty-eight filtering blebs of 28 patients with open-angle glaucoma were measured horizontally and vertically in the sub-Tenon space with AS-OCT after ab externo SIBS microshunt implantation with MMC. The intraocular pressure (IOP) was monitored simultaneously at each visit. Maturation of and morphological changes in the blebs and correlations with the IOP were recorded. RESULTS: The average median preoperative IOP of 20.7 (range, 12-30) mmHg decreased to 8.5 (range, 4-17), 8.9 (range, 5-17), 10.4 (range, 8-16) and 10.9 (range, 9-15) mmHg at 24 hr, 1 week, 1 month and 3 months, respectively (p < 0.001). A multiform morphology on AS-OCT prevailed at all time points, with a 3.5% rate of a uniform bleb morphology at the first week. The horizontal and vertical diameters of the blebs increased from baseline to the third month. The horizontal expansion (406 ± 127 μm on day 7, p = 0.04, 712 ± 211 μm on day 30, p = 0.02 and 952 ± 218 μm on day 90, p < 0.001) was greater than the vertical expansion (16 ± 18 μm, p = 0.3 on day 1, 63 ± 27 μm, p = 0.02 on day 30 and 137 ± 34 μm, p < 0.001 on day 90) without correlation with the IOP (r = -0.3, p = 0.2). CONCLUSION: Anterior-segment OCT (AS-OCT) of the filtering blebs formed after ab externo SIBS microshunt implantation showed progressive horizontal and vertical expansion of the blebs in the sub-Tenon space, with a significant peak at the first month not significantly correlated with the decrease in the IOP.