advertisement

Topcon

Abstract #98909 Published in IGR 22-4

Clinical outcomes of a beveled tip, ultra-high speed, 25-gauge pars plana vitrectomy system

Uy HS; Cabahug VLO; Artiaga JCM; Chan PS; Famadico JT
BMC Ophthalmology 2022; 22: 93


OBJECTIVE: To report the clinical outcomes of a 25-gauge, beveled-tip, 10,000 cuts-per-minute (cpm) microincisional vitrectomy surgery (MIVS) system. METHODS: Prospective case series of eyes undergoing primary pars plana vitrectomy (PPV) for common vitreoretinal indications. Main outcome measures were: rate of achieving surgical objectives, operative times, number of surgical steps, use of ancillary instruments, corrected distance visual acuity (CDVA), and adverse events (AE). RESULTS: The surgical objectives were achieved in all eyes. Mean total operative time (TOT), core, shave and total vitrectomy times were 1891 ± 890, 204 ± 120, 330 ± 320, 534 ± 389 s, respectively. Mean number of surgical steps was 4.3 ± 1.5. Mean number of ancillary instruments used was 4.5 ± 1.9. Mean CDVA improved by 0.53 ± 0.56 logMAR units (P < 0.001) 3 months postoperatively. AE included elevated IOP (8%), hypotony (6%), and re-detachment (2%). Majority (82%) had no postoperative discomfort. The number of surgical steps demonstrated a positive correlation with TOT (p < 0.05), number of ancillary instruments used (p < 0.05), and postoperative Day 1 IOP (p < 0.05). The number of times ancillary instrumentation was used demonstrated a positive correlation with TOT (p < 0.05). CONCLUSION: Beveled-tip, 10,000 cpm MIVS system effectively and safely performs common VR procedures of varying complexity and may reduce operative times and use of ancillary instrumentation.

Full article

Classification:

15 Miscellaneous



Issue 22-4

Change Issue


advertisement

Oculus