PURPOSE: We describe a case of conjunctival erosion following a PRESERFLO® MicroShunt procedure, and the subsequent revision surgery. OBSERVATIONS: Conjunctival erosion was noted overlying the MicroShunt implant at postoperative week 11, 8 weeks following a bleb needling procedure for bleb encapsulation. A brisk leak was observed at the site of erosion. The patient underwent a subsequent revision procedure with repositioning of the MicroShunt implant and mitomycin C (MMC) application. CONCLUSION AND IMPORTANCE: Conjunctival erosion may be a relatively rare but important complication following MicroShunt surgery and may arise from a variety of risk factors. Extra care should be taken during bleb needling in the context of MicroShunt, and needling should be directed posteriorly, beyond the distal tip of the MicroShunt.
St Thomas' Hospital, Department of Ophthalmology, Westminster Bridge Road, London, United Kingdom.