PURPOSE: To determine outflow resistance of the Baerveldt glaucoma implant using different tube configurations. METHODS: Outflow resistance of 6 tube configurations (C1- C6) of Baerveldt implants was measured under conditions of constant pressure perfusion. Pressures ranged from 2 to 55 mmHg. Venting slits were created using a 7-0 Vicryl, spatulated suture-needle. Seton tubes were occluded by threading a retrograde suture approximately 1.5 cm into the lumen. RESULTS: At pressures between 2 and 55 mmHg, mean outflow resistance of the normally configured seton (ie, open tube; C1) was 0.41 (± 0.6) mmHg/μL/min. Resistance was unchanged (mean 0.41 (± 0.4) mmHg/μL/min) by the addition of 4 venting slits (C2) to the seton tube. Occlusion of the open seton tube with a 3-0 Supramid suture (C3) significantly increased (P < 0.001) mean outflow resistance to 14.99 (± 0.6) mmHg/μL/min. Occlusion of the tube with a 4-0 Supramid suture (C4) significantly increased (P < 0.001) mean outflow resistance to 1.09 (± 0.5) mmHg/μL/min. In implants where tubes were occluded with a 3-0 Supramid suture, the addition of venting slits (C5) significantly decreased (P = 0.038) mean outflow resistance to 8.98 (± 0.4) mmHg/μL/min. In tubes occluded with a 4-0 Supramid suture, the addition of venting slits (C6) decreased mean outflow resistance to 0.98 (± 0.6) mmHg/μL/min. CONCLUSIONS: Although these results cannot be directly correlated to the clinical setting, they do show that outflow resistance can be modified at the time of surgery by changing tube configuration of the Baerveldt glaucoma implant. Configuration C5 (3-0 Supramid with venting slits) closely approximates the outflow rate in the normal intraocular pressure range.
Dr. R.R. Breckenridge, Hewitt Laboratory of the Ola B. Williams Glaucoma Center, Department of Ophthalmology, Medical University of South Carolina, Charleston, South Carolina, USA
12.8.2 With tube implant or other drainage devices (Part of: 12 Surgical treatment > 12.8 Filtering surgery)