advertisement
The 20-MHz ultrasound probe was compared with ultrasound biomicroscopy to determine its usefulness in imaging various glaucomatous conditions. Ten patients with glaucoma underwent anterior segment imaging with both the 20-MHz probe, which attaches to the I3 B-scan (Innovative Imaging Inc., Sacramento, CA), and the Ultrasonic BioMicroscope (UBM; Paradigm Medical Industries, Salt Lake City, UT). All pathology was easily demonstrable using the 20-MHz probe, in one case showing a retinal detachment not seen with the UBM. However, anterior findings such as fluid in the suprachoroidal space and sclerostomy sites in postoperative trabeculectomy cases were more difficult to view with the 20-MHz probe. The use of coupling enhanced the quality of the latter images. The 20-MHz ultrasound probe may be a viable aid in diagnosis and follow-up of certain glaucomatous conditions, and the use of a coupling device enhances its images.
Dr. P.J. Harasymowycz , William and Ana Goldberg Glaucoma Service, Wills Eye Hospital, Philadelphia, USA
6.12 Ultrasonography and ultrasound biomicroscopy (Part of: 6 Clinical examination methods)