A novel 8-mm Schlemm's canal scaffold reduces outflow resistance in a human anterior segment perfusion model

Gulati V; Fan S; Hays CL; Samuelson TW; Ahmed II; Toris CB
Investigative Ophthalmology and Visual Science 2013; 54: 1698-1704


PURPOSE: To study the effect on outflow facility and outflow resistance of a nitinol microstent implanted into Schlemm's canal. METHODS: Using a constant pressure perfusion method, outflow facility and outflow resistance were measured in 26 pairs of dissected anterior segments from donated human eyes. Measurements were made at perfusion pressures of 10, 20, 30 and 40 mm Hg. The Hydrus Microstent was placed in Schlemm's canal of one eye and the contralateral eye underwent a sham procedure. Outflow facility and outflow resistance were measured again after the microstent implantation or sham procedure. RESULTS: The Hydrus Microstent significantly increased outflow facility from 0.33 ± 0.17 μL/min/mm Hg to 0.52 ± 0.19 μL/min/mm Hg (P < 0.001). Outflow resistance was significantly reduced from 4.38 ± 3.03 mm Hg/μL/min at baseline to 2.34 ± 1.04 mm Hg/μL/min (P < 0.001) with the microstent. There was a linear correlation between outflow resistance at baseline and decrease in outflow resistance with the microstent (R(2) = 0.89, P < 0.0001). CONCLUSIONS: The increase in outflow facility and decrease in resistance supports the potential use of the Hydrus Microstent as a surgical option to reduce intraocular pressure (IOP). The IOP-lowering effect may be higher in eyes with higher outflow resistance (and IOP) as compared with eyes with lower outflow resistance (and IOP).

Department of Ophthalmology and Visual Sciences, University of Nebraska Medical Center, Omaha, NE 68198, USA.

Full article

Classification:

12.8.2 With tube implant or other drainage devices (Part of: 12 Surgical treatment > 12.8 Filtering surgery)



Issue 15-1

Change Issue


advertisement

Oculus