advertisement

Topcon

Abstract #11324 Published in IGR 6-3

Reliability of intraocular pressure by Tono-Pen XL over amniotic membrane patch in human

Yoshita T; Kobayashi A; Takahashi M; Sugiyama K
Journal of Glaucoma 2004; 13: 413-416


PURPOSE: To determine the reliability of Tono-Pen XL in measuring intraocular pressure (IOP) over the amniotic membrane patch in human eyes. METHOD: Ten healthy volunteers (nine males and one female) participated in this study. After topical anesthesia, IOP was measured using the Tono-Pen XL before and after the application of a single-layer amniotic membrane patch in the right eye and double-layer amniotic membrane patch in the left eye. RESULTS: We found no statistical difference between the IOP measured with and without a single layer amniotic membrane (P = 0.808). Nor did we find any statistical difference between the IOP measured with and without double-layer amniotic membrane patch (P = 0.813). The average thickness of the single- and the double-layer amniotic membrane patch measured by pachymetry was 131.9 ± 31.4 μm (mean ± SD) and was 246.1 ± 69.3 μm (mean ± SD), respectively. CONCLUSIONS: This study showed that it is reliable to measure IOP by Tono-Pen XL over single or a double layer of amniotic membrane patch on healthy human eyes. Further studies are required to determine the accuracy of IOP measurement over amniotic membrane patch on diseased corneas.

Dr. T. Yoshita, Department of Ophthalmology, Kanazawa University Graduate School of Medical Science, Kanazawa, Japan


Classification:

6.1 Intraocular pressure measurement; factors affecting IOP (Part of: 6 Clinical examination methods)



Issue 6-3

Change Issue


advertisement

WGA Rescources