advertisement

WGC-2027

Abstract #127497 Published in IGR 25-3

Surgical technique to prevent malposition of the Hydrus glaucoma microstent (The "gap sign")

Andrew NH; Andrew NH; Habib M
Indian Journal of Ophthalmology 2025; 73: S515-S516


The key challenge of implanting the hydrus glaucoma microstent is avoiding the tip from exiting Schlemm's canal and diving posteriorly toward the iris root. Here, we propose a novel theory as to why the hydrus stent tends to dive posteriorly and we describe surgical steps to prevent this from occurring. These steps include making a corneal incision oriented tangential to the angle, having a flat angle of attack to Schlemm's canal, and relaxing the hand before advancing the stent. Observing a gap between the stent and the back wall of the hydrus injector ("gap sign") indicates that the stent is striking the back wall of Schlemm's canal and is likely to dive posteriorly unless the surgeon's hand position is adjusted.

Sight Specialists, Gold Coast, Australia.

Full article

Classification:

15 Miscellaneous



Issue 25-3

Change Issue


advertisement

WGA Rescources