Is pneumotonometry still reliable?

Bresson-Dumont H; Lehoux A; Kponoume C; Santiago P-Y 
Journal Français d'Ophtalmologie 2007; 30: 3S66-3S72


Noncontact tonometer (NCT) is an applanation tonometry, as is Goldmann tonometry. Numerous studies have shown the reliability of this technique but problems still exist. METHOD: We measured intraocular pressure (IOP) in 340 patients (668 eyes) with NCT and Goldmann tonometer (GAT). To evaluate the influence of central corneal thickness (CCT), we classified the patients according to their CCT (< 500 μm; between 520 and 580 μm, and > 600 μm). RESULTS: The mean difference between the two IOP measurements was -1.4 mmHg ± 2.67 (-12 to +7), p < 0,0001 for the thinner corneas, +0.13 mmHg ± 2.67 (-8 to +8), p = 0.445 for the regular corneas, and + 3.04 mmHg ± 3.2 (-7 to +13) p < 0.0001 for the thicker corneas. CCT influences NCT and GAT, but the difference between the two techniques is higher for thicker CCTs, higher IOPs, and in younger patients. CONCLUSION: NCT is good for glaucoma screening but is less reliable than GAT for glaucoma follow-up if CCT and IOP are far from regular values. LA: French


Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)