advertisement

WGA Rescources

Abstract #24310 Published in IGR 11-3

Comparison between dynamic contour tonometry and Goldmann applanation tonometry

Lee J; Lee CH; Choi J; Yoon SY; Sung KR; Park SB; Kook MS
Korean Journal of Ophthalmology 2009; 23: 27-31


PURPOSE: To compare the intraocular pressures (IOPs) measured by dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT), and to investigate the association of IOPs on eyes of varying central corneal thickness (CCT). METHODS: In this prospective study, 451 eyes of 233 subjects were enrolled. IOPs were measured by GAT and DCT. CCT was measured three times and the average was calculated. Each eye was classified into one of three groups according to CCT: low CCT (group A, CCT<520 µm, n=146); normal CCT (group B, 520 µm550 µm, n=142). In each group, we investigated the association of CCT with IOP measurement by GAT and DCT. RESULTS: The IOPs measured by GAT and DCT were significantly associated for all eyes (R=0.853, p<0.001, Pearson correlation). CCT was related with both IOP measurement by GAT and DCT with statistical significance (mixed effect model, p<0.001). However, subgroup analysis showed that CCT affected IOP measured by GAT for groups B and C, whereas it affected IOP measured by DCT only for group C. CONCLUSIONS: IOP measured by DCT was not affected by CCT in eyes with low to normal CCT, whereas this measurement was affected in eyes of high CCT range. CCT may have less effect on IOP measurements using DCT than those obtained by GAT, within a specified range of CCT.

Dr. J. Lee, Department of Ophthalmology, College of Medicine, University of Ulsan, Asan Medical Center, Seoul, Korea


Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)



Issue 11-3

Change Issue


advertisement

Oculus