A 62-year-old woman with neovascular glaucoma underwent filtering surgery with the Ahmed glaucoma valve (New World Medical, Inc., Rancho Cucamonga, CA). Intraoperative hypotony secondary to a defective valve was noted and managed by excising the valve mechanism. The implant was then treated as if it was non-valved by placement of an obturator suture. The postoperative course was unremarkable for a non-valved implant. This case demonstrates how intraoperative hypotony secondary to overfiltration of an Ahmed glaucoma valve implant can be successfully managed by converting to a non-valved implant.
Dr. N.P. Bell, Department of Ophthalmology and Visual Science, University of Texas Medical School at Houston, 6411 Fannin Street, Houston, TX 77030, USA
12.8.2 With tube implant or other drainage devices (Part of: 12 Surgical treatment > 12.8 Filtering surgery)